Some problems will happen in the using of the microscope, you could solve them according the following list. If you cannot solve the problem after checking the entire list, please contact us for assistance.
| |
PROBLEM |
REASON FOR PROBLEM |
SOLUTION |
| *** Optical Part *** |
| 1. |
Illumination is opening, but the field of view is dark. |
The poor contact exists in the lamp house and the illumination system. |
Ensure the contact pin and the lamp holder pin work well. |
| The lamp bulb spoils. |
change a new bulb. |
| The brightness adjustment knob is set too dark |
Adjust the knob in a proper position |
| No use the appointed lamp bulb |
use the specified halogen Lamp 6V30W |
| 2. |
The edge of the field of view has shadow or the brightness is asymmetry |
The nosepiece is not in the located position |
Adjust it into the located position |
| The surface of the lamp become black |
Change a new lamp bulb |
| The surface of the lens is moldy or has contaminant |
Clean the lens |
| 3. |
Find dust and stain in the field of view |
There are stains on the specimen |
Change the specimen |
| There are stains on the eyepiece |
Clean the eyepiece |
| 4. |
The image is defocus\low-resolution |
The objective damage |
Mend and correct the objective (send to factory for overhauling) |
| The lens of the objective and eyepiece is moldy or have contaminant |
Do cleaning |
| The opening of Aperture diaphragm and field diaphragm is not proper, and too much astigmatism. |
Change the opening of the aperture diaphragm and field diaphragm |
| Fine focus system is broken |
Examine and repair the fine focus system (send to factory for overhauling) |
| The objective is not in the center of the light path |
Turn the nosepiece to the located position |
| 5. |
The image focus surface incline (one side is clear and the other side is faint) |
The illumination light incline serious |
Adjust the filament position ,let the light distributing of the field of view become symmetrical and bright |
| The specimen don’t correctly place |
Put the specimen on the right position |
| The nosepiece is not in the located position |
Turn the nosepiece in the required position |
6. |
The eyes are uncomfortable, the left and right fields of view is not coincided |
The interpupillar distance is not correct |
Adjust the interpupillar distance correctly |
| The diopter is not right |
Adjust the diopter according your sight |
| Can’t adapt to binocular observation |
When look into the objective, do not stare at the specimen but at the whole field of view, or move the eyes away to see other things, then back into the objective |